CHAPTER 9

ULTRASONIC TESTING OF ALUMINO THERMIT WELDED JOINTS

(Chapter revised vide ACS  No.6 of 1999)

9.1 Scope : This procedure covers the requirement of ultrasonic testing of alumino thermit (AT) welded rail joints immediately after execution of the weld and also for periodic examination.

9.2 General conditions of test

9.2.1 Surface preparation: After execution of the AT weld, the welded zone shall be dressed properly to facilitate placement of probes and to avoid incidence of spurious signals on the CRT. The rail table shall be  dressed to obtain reasonably flat and smooth surface. The flange and the web up to a distance of 200 mm on either side of the weld collar shall be thoroughly cleaned with a wire brush to ensure freedom from dust, dirt, surface unevenness etc. (ACS No.6 of 1999)

9.2.2 Couplant : Water/oil /soft grease shall be used as couplant. (ACS No.6 of 1999)

9.2.3 Sensitivity : The equipment sensitivity shall be set for normal, 70and 70, 1.25 MHz  probes in accordance with the procedure laid down in para 9.4. The sensitivity so adjusted shall be considered as normal gain sensitivity and shall be utilised during ATW testing. The sensitivity level shall not be altered during the course of testing. (Para 9.2.3 revised vide ACS  No.10 of  2002 by superceding ACS No.6 of 1999)

9.3 Apparatus required

9.3.1 Equipment: Any model of RDSO approved rail tester shall be considered suitable for testing of AT welded rail joints.

9.3.2 Probes: During ultrasonic examination of AT welded joints, the following probes shall be utilized   :

a.     Normal (0), 4 MHz     -Double crystal fitted in trolley.

b.     70, 2 MHz                 -Double crystal fitted in trolley.

c.       70, 2.0 MHz             -Single crystal for hand probing

9.3.3 Cable : One co-axial cable of suitable length for connecting 70, 2.0 M.Hz probe to flaw detector shall be used. The length should not exceed more than 5 m. (Paras 9.3.2 and 9.3.3 revised vide ACS  No.10 of  2002 by superceding ACS No.6 of 1999)

 

9.4   Sensitivity setting procedure

 

9.4.1 Calibration : The equipment should be set for a depth range of 300 mm (longitudinal wave) by adjusting depth control knob suitably. Each division, therefore, shall correspond to 30 mm. (longitudinal waves) (Para 9.4.1 revised vide ACS  No.10 of  2002 superceding ACS No.6 of 1999)

 

9.4.2 Test rail: The sensitivity of the ultrasonic equipment shall be set with the help of a standard AT welded rail piece of 1.5 m length having a simulated flaw at standard locations as shown in Fig. 17.

 

9.4.3 Alignment of probes : The alignment of normal and 70 probes fitted with the trolley may be checked by placing the rail tester on the test rail using water/oil as couplant and ensuring that the probes travel along the vertical axis of the rail.

 

9.4.4 Sensitivity setting

 

9.4.4.1 Place the trolley on the test rails shown in fig. 17. Keep the switches of all the probes in off positions and turn the potentiometer knobs of all the probes to 50% of their highest working range.

 

9.4.4.2 Switch on only 70forward probe and move the equipment towards the drilled hole of 3 mm dia in the rail head. When the probe is just in the reflecting range, a pulse corresponding to the hole shall appear on the screen which during onward travelling shall show higher amplitude. The pulse shall appear moving from right to left The equipment should be progressively moved forward till maximum height of the pulse is obtained. At this location, the height of the pulse shall be adjusted to 60% of full screen height by suitable manipulation of the gain knob.

 

9.4.4.3 The forward probe shall be switched off and the 70% backward probe shall now be switched on. In this case a flaw signal shall appear moving from left to right. The signal height in this position shall also be adjusted to 60% of full screen height. This can be accomplished through suitable manipulation of relevant potentiometer.

 

9.4.4.4 The sensitivity setting for the normal probe has to be done while keeping all other probes in off position. Switch on only the normal probe and bring it above 3 mm dia hole at head of the test rail. Manipulate the potentiometer control knob to obtain echo height of 60% of full screen height at 1.0 division horizontal scale.

 

9.4.4.5 70, 2.0 MHz  probe shall be connected to the socket available in the ultrasonic equipment The selectors switch may be set to single crystal mode. Move the probe towards the 3 mm dia hole drilled at the middle of flange through the AT weld and manipulate knobs to obtain a   60% full screen height on the CRT.

 

9.5     Criteria for defect classification for initial testing

 

9.5.1  Till such time more exhaustive studies are carried for strengthening the initial testing of AT welds, any flaw signal obtained by normal probe in the head of 40% height or more shall be treated as a defective AT welded joint. Any flaw signal  of 10% height  or more obtained from the normal probe either from the web or the foot location shall also be a cause for rejection of the AT weld. (Paras 9.4.4.5 and 9.5.1 revised vide ACS  No.10 of  2002 by superceding ACS No.6 of 1999)

 

9.5.2 In the case of lack of fusion, inclusions, blow holes etc, in the rail head, moving signal shall be obtained while testing with 70probe. The position of onset of the signal and its corresponding range on the horizontal screen as well as their maximum amplitude shall be recorded.

A welded joint showing the moving signal of 40% or more of the full screen height shall be considered as a defective welded joint.

 

9.5.3    70, 2.0 MHz  probe shall be placed on the flange at a distance of 180 mm  corresponding to position 'L' in Fig. 18 such that ultrasonic waves are directed towards the weld. The probe shall thereafter be moved slowly in a zigzag pattern towards the weld. A welded joint showing a flaw echo of 40% vertical height or more with the stipulated gain setting shall be treated as a defective welded joint. Similar testing shall be carried out from 'C' and 'U' region as shown in Fig. 18.   In these cases also the criteria for rejection shall remain the same. (Para 9.5.3  revised vide ACS  No.10 of  2002 superceding ACS No.6 of 1999)

 

9.6 - A thermit welding done in-situ shall be joggled fish plated with four clamps and supported on wooden blocks till tested as good by USFD" (ACS No. 6 of 1999)

 

9.6.1 The defective joints based on the criteria mentioned at para 9.5 shall not be allowed to remain in service and these joints shall be cropped, re-welded and tested again. This execution shall be done by the contractor free of cost. The re-welded joints shall be scanned ultrasonically again with the same set of acceptance criteria to ensure freedom from any harmful detects.

 

 

 

 

 

9.7 Periodic testing of welds in service

 

9.7.1 Testing with 0 and 70probes : During on track periodic testing of AT welded rail joints gain setting shall be done in respect of 0 and 70angle probes on standard test rail, (need based/conventional as the case may be) as applicable for rail testing (Refer Fig. 3 and 4).

 

9.7.1.1  The testing of weld using 0 and 70probe shall be conducted with the help of rail test trolley during routine rail testing.

 

9.7.2 Testing with 70, 2.0 MHz  probe (Flange testing): Gain setting of the USFD equipment during testing shall be such that a flaw peak height of 80% of full screen height is obtained from 3 mm dia hole at the flange through the AT weld (Fig. 17).

 

9.7.2.1  The testing of the flange using 70, 2.0 MHz  probe shall be conducted in the same manner as outlined for testing of new AT welds (Para 9.4.4.5). (Para 9.7.2 and 9.7.2.1 revised vide ACS  No.10 of  2002 by superceding ACS No.6 of 1999)

 

9.8 Criteria for classification of defective joints based on 0and 70probe shall be as per Annexure IV for testing with conventional criteria and as per Annexure VII for need based criteria.

 

9.9  Criteria for classification of defective joints in the flange location tested by 70, 2.0 MHz  probe shall be as per Annexure V for conventional testing and as per Annexure VII for need based concept of USFD testing. (Para 9.9 revised vide ACS  No.10 of  2002 by superceding ACS No.6 of 1999)

 

9.10.     Procedure for detection of half noon cracks in AT welds.

 

9.10.1 Preamble: Half moon cracks are found to be present in the rail foot centre of AT welds and are oriented transversely.  In AT welds these defects are associated with fins formed during AT welding.  For avoiding such defects and consequent failure in service, the following procedure shall be followed :

 

9.10.2          Procedure: All AT welds in service are required to be examined visually and through ultrasonic examination for detection of cracks emanating from rail foot.  Visual examination can be carried out by PWI/USFD or sectional PWI and USFD examination shall be carried out by PWI/USFD.  The procedure for visual and USFD examination shall be as under.

 

9.10.3          Visual Examination:  All the welded joints shall be cleaned and examined carefully to detect any visible defects like cracks, blow holes, shrinkage, mismatch, surface finish (smooth surface required) etc.  Any joint which shows visible defect shall be declared defective.

 

                   The bottom of the joint shall be checked by feeling with fingers as well as inspected with the help of a mirror for presence of fins at the parting line of the mould.  If fin is observed in any joint, the joint shall be declared defective.”

 

9.10.4.                      Ultrasonic Examination

 

9.10.4.1                Scope-This procedure covers the requirements of ultrasonic examination of Alumino-Thermic joints for detection of half moon cracks emanating transversely from the centre of the rail foot.

 

9.10.4.2                Apparatus required:

 

9.10.4.2.1        Equipment-Any RDSO approved equipment meeting the requirements of   RDSO specification. 

 

9.10.4.2.2    Probes:      The following probes shall be used for Ultrasonic Examination of AT welded joints.

 

(i)                Normal (0), 2 MHz 18mm PZT double crystal probe for calibration.

(ii)             45 /2MHz, 20mm x 20mm PZT crystal forward and backward probe housed in one assembly or single 45 /2 MHz, 20mm x 20mm PZT crystal probe.

 

(iii)           70 /2 MHz, 8mmx8mm, PZT crystal probe-1 Nos.

 

 

9.10.4.2.3    Cables:      Coaxial cable for each probe shall be used.  The length shall be approximately 2 metre.

 

9.10.4.2.4    Couplant:  Soft grease/Oil/water shall be used as couplant.

 

9.10.4.3       General condition of test:   The welded zone of the joint shall be dressed on all sides properly to facilitate placement of probe and to avoid incidence of spurious signals on the CRT. The head surface shall be dressed to obtain reasonably flat and smooth surface.  The outer faces of the flange shall also be clean and smooth.

 

9.10.4.4.      Standard Test Piece: Standard A.T. welded rail piece of 1.5 metre length with AT joint in the middle having simulated flaw as shown in fig.36.

 

9.10.4.5                 45º forward and Backward probe

 

9.10.4.5.1    Range calibration:      The equipment shall be set for depth range of 500mm for longitudinal wave using 0 probe.

 

9.10.4.5.2    Sensitivity setting:     Place 45 /2MHz (F&B) probe or 45 / 2MHz probe on the rail head surface at a distance equal to height of rail from the centre of the AT weld Move the probe 40mm either side of this position to pick up half moon crack at lower fins on zone of the flange.  This exercise shall be carried out twice, once from each side of the weld and signal from simulated flaw as shown in fig.  Should appear at a distance approx 400mm for 52 kg rail.  This distance will vary with respect to rail section height.  The signal so obtained shall be adjusted to 60% of full screen height by manipulating the gain control knob.

 

9.10.5          Test Procedure: The probe 45 /2MHz (F&B) probe or 45 /2MHz single probe shall be placed on the railhead of AT weld under test with same sensitivity as per para 9.10.4.5.2.  This testing technique will scan the bottom of the flange weld in the central zone.  If 45 2MHz single probe is used the scanning shall also be done from one side of railhead of the weld with beam directing towards the web foot region of the weld.  45 / (F&B) probe shall be moved on the railhead table crossing the weld region on the other side as shown in fig.37.

 

9.10.6         70 /2MHz, size 8mm x 8mm probe.  This technique shall be used when  due to presence of hole in the web; the above procedure is no possible to adopt.

 

9.10.6.1       Range calibration:      The equipment shall be set for a depth range of 300 mm (longitudinal wave) by manipulating the depth control knob suitably.

 

9.10.6.2.      Sensitivity Setting:     One 70/2 MHz, size 8mmx8mm probe shall be connected to the unit by means of one coaxial cable.  The selector switch shall be in T+R (single crystal) mode.  Apply the couplant and place the probe at a distance of 100 mm from edge of the weld near to root of the flange with beam direction towards simulated flaw.  Move the probe slightly in zigzag manner to get reflection from simulated flaw.  Set the height of signal to 60% of full screen height.  The gain level will be utilized during testing.

 

9.10.6.3       Test Procedure: 70 probe shall be placed on the root of the flange at a suitable distance (100mm approx) with beam directing toward web-foot junction and scanning shall be carried out as per para 9.10.6.2.  This exercise shall be carried from all the four sides of rail flange and any moving signal on the screen shall be recorded.

 

                   The method is not capable of detecting all half moon defects due to its position of probing.

 

9.10.7          Defect Classification: Any flaw signal of height 10% or more  than of the full screen height shall be considered as defective AT weld in both the technique applied for 45 and 70º probes. and shall be classified as DFW.

(New para 9.10 introduced vide ACS No. 12 of 2002)

(Underlined extension done in para 9.10.7 vide ACS No.13 of 2003)